Description
It is well known that the old 1GHz current target described in previous editions of IEC 61000-4-2 had limitations that prevented the user from seeing high frequency ringing and noise that might exists. Now that higher bandwidth oscilloscopes are now readily available, the improved 4 GHz ESD current calibration kit becomes necessary – otherwise measurements are still limited by the older target design.
A4001 ESD Current Target Calibration Kit (designed by the original designer of the ESD Current Target in standard committee, David Pommerenke) is a DC-4 GHz wide-band ESD Target for ESD simulators calibration according to the latest IEC 61000-4-2 and ISO 10605 standards. The attenuator provides return loss and the cable is used as the connection to the oscilloscope. The frequency response specifications are taken with the complete assembly.
A4001 ESD Current Target Calibration Kit is comparable to Haefely PET 4000, Teseq MD 103, Noiseken 06-00067A, and EM Test CTR2.
ESDEMC has been the ESD Current Target supplier for Haefely Hipotronics, Thermo Fisher Scientific, ESDGuns.com, etc…
Applications:
ESD Current Measurement
ESD Simulator Discharge Current Calibration
A4001-Kit-N-20+6 Specifications:
ESD Target Surface Finish | Gold Plated, never oxidized or rusted |
Measurement Bandwidth | DC – 4 GHz < ± 0.5 dB up to 1GHz < ± 1.2 dB up to 4GHz |
Input Resistance | 2.08 Ohm (± 1%) (2.00 Ohm in parallel with 50 Ohm load) |
Max. IEC Pulse Level | Tested up to ± 30kV |
Voltage Non-linearity | <2% up to ± 20 kV < 5% up to ± 30kV |
Coaxial Cable Type | RG400, Double Shielded |
Coaxial Cable Length | 0.8 m |
Coaxial Cable Connector | SMA – N Type |
Transfer impedance | 5A/V with default 20dB attenuator |
Meets Standards | IEC 61000-4-2 ED. 2.0 B:2008 ISO 10605:2008 |
Default Configuration:
A4001-N | Gold Plated Wide-band ESD Target with N Type Connector |
AT2WN-20 | N Type Attenuator, DC-18 GHz, 2W , 20 dB |
RG4-NS-8 | RG400 Coaxial Cable, N-SMA Connectors, 0.8 meter |
A4001-SMA2BNC | High Precision SMA-BNC Adapter |
A4001-CAL | NIST Traceable Factory Calibration Report |
Optional Accessories:
A4001-SMA | Gold Plated Wide-band ESD Target with SMA Connector |
AT2WN-6 | N Type Attenuator, DC-18 GHz, 2W, 6 dB (Use in addition with 20 dB ATT for up to 30 kV calibration) |
AT2WN-30 | N Type Attenuator, DC-18 GHz, 2W, 30 dB (Use instead of default 20 dB ATT for up to 30 kV ESD calibration) |
AT2WS-20 | SMA Type Attenuator, DC-18 GHz, 2W, 20 dB |
RG4-SS-8 | RG400 Coaxial Cable, SMA-SMA Connectors, 0.8 meter |
A4001-Repair | ESD target repair (will work with other brands ESD Targets) |
A4001-ReCal | Re-Calibration of ESD Target (NIST Traceable) |
A4001-TP | ESD Current Calibration Shielding Plane |
A4001-TSC | ESD Current Calibration Shielding Cabinet with EMI Filter |